D.2.2 Identify the smallest capacity load cells in each group to be tested, according to R 60-2, 2.4:

Class

Group

<—- lowest

—-> highest

C3

C6

1

50

B10

In this example, select and identify:

C6 — 50 kg (full evaluation test required)

B10 — 500 kg (full evaluation test required)

Although load cell C3 — 100 kg is the smallest capacity in its group, its capacity falls within the range of other selected load cells having better metrological characteristics. Therefore, it is not selected.