D.2.2 Identify the smallest capacity load cells in each group to be tested, according to R 60-2, 2.4:
Class Group | | <—- lowest | | —-> highest | |||||||
C3 | | | | | | | | | |||
C6 1 | | 50 | | | | ||||||
B10 | | | | | |||||||
In this example, select and identify:
C6 — 50 kg (full evaluation test required)
B10 — 500 kg (full evaluation test required)
Although load cell C3 — 100 kg is the smallest capacity in its group, its capacity falls within the range of other selected load cells having better metrological characteristics. Therefore, it is not selected.